BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

47

4.2.7

Output - AC Voltage

CALibrate[:SOURce]:VOLTage[:AC]:LRANge[:FSCale]

This command will result in the calibration coefficent being calculated for the AC full scale
output voltage at the low voltage range.

Command Syntax

CALibrate:VOLTage:LRANge

Parameters

(programmed output voltage)

Examples CAL:VOLT:LRAN

120

Query Syntax

CALibrate:VOLTage:LRANge?

Returned Parameters

(cal coefficient value range 0.9 to 1.1)

Related Commands

CAL:VOLT:LRAN:ZERO CAL:VOLT:LRAN:HFR

CALibrate[:SOURce]:VOLTage[:AC]:LRANge:ZERO

This command will set the calibration coefficent for the output voltage offset at the low
voltage range.

Command Syntax

CALibrate:VOLTage:LRANge:ZERO

Parameters

(a value between -127 and +128)

Examples

CAL:VOLT:LRAN:ZERO +10

Query Syntax

CALibrate:VOLTage:LRANge:ZERO?

Returned Parameters

(value range -127 to +128)

Related Commands

CAL:VOLT:LRAN CAL:VOLT:LRAN:HFR

CALibrate[:SOURce]:VOLTage[:AC]:LRANge:HFRequency

This command will set the calibration coefficent for the AC full scale output voltage at the low
voltage range and at high output frequency

Command Syntax

CALibrate:VOLTage:LRANge:HFRequency

Parameters

(a value between -127 and +128)

Examples CAL:VOLT:LRAN:HFRequency

+10

Query Syntax

CALibrate:VOLTage:LRANge:HFRequency?

Returned Parameters

(value range -127 to +128)

Related Commands

CAL:VOLT:LRAN CAL:VOLT:LRAN:ZERO

CALibrate[:SOURce]:VOLTage[:AC]:HRANge[:FSCale]

This command will result in the calibration coefficent being calculated for the AC full scale
output voltage at the high voltage range.

Command Syntax

CALibrate:VOLTage:HRANge

Parameters

(programmed output voltage)

Examples CAL:VOLT:HRAN

-2

Query Syntax

CALibrate:VOLTage:HRANge?

Returned Parameters

< (cal coefficient value range 0.9 to 1.1)

Related Commands

CAL:VOLT:HRAN:ZERO CAL:VOLT:HRAN:HFR