SENSe:SWEep:TINTerval
This command and its query format allow setting the time period between samples when
voltage and current digitization is controlled by the acquire trigger sequence. The
query response of the sample period query will be:
MX Series I: 25.6 to 256
μsec for single phase and 76.8 to 768 μsec for three phase AC/DC
source configurations respectively. The value of TINT must be set in increments of 25 or 75
μsec. The controller will round to the nearest available multiple of 25.6 or 76.8 μsec.
MX Series II and RS Series: 10.4 to 104
μsec for single phase and 31.2 to 312 μsec for
three phase AC/DC source configurations respectively. The value of TINT must be set in
increments of 10.4 or 31.2
μsec. The controller will round to the nearest available multiple of
10.4 or 31.2
μsec.
Command Syntax
SENSe:SWEep:TINTerval
Parameters
MX Series I:
25.6 to 256 for single phase configuration
76.8 to 768 for three phase configuration
MX/RS Series II:10.4 to 104 for single phase configuration
31.2 to 312 for three phase configuration
*RST Value
MX Series I:
25.6 or 76.8
MX/RS Series II:10.4 or 31.2
Examples SENS:SWE:TINT-150
Query Syntax
SENSe:SWEep:TINTerval?
Returned Parameters
Related Commands
SENS:SWE:OFFS
MEAS:ARR