BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

109

VOLTage:MODE

Phase Selectable

This command determines how the output voltage is controlled during a triggered output
transient. The choices are:

FIXed

The voltage is unaffected by a triggered output transient.

STEP

The voltage is programmed to the value set by VOLTage:TRIGgered when
a triggered transient occurs.

PULSe

The voltage is changed to the value set by VOLTage:TRIGgered for a
duration determined by the pulse commands.

LIST

The voltage is controlled by the voltage list when a triggered transient
occurs.

Command Syntax

[SOURce:]VOLTage:MODE

Parameters

FIXed | STEP | PULSe | LIST

*RST Value

FIX

Examples

VOLT:MODE LIST

VOLT:MODE FIX

Query Syntax

[SOURce:]VOLTage:MODE?

Returned Parameters

Related Commands

VOLT:TRG

VOLT

VOLTage:HIGH

This command programs the maximum rms voltage that the power source will accept. The
maximum value will be the lower of this value or the voltage range.

Command Syntax

[SOURce:]VOLTage:HIGH

Parameters

0 to maximum voltage range specified by the LIMit:VOLTage
command

Unit

V (rms voltage)

*RST Value

no change

Examples VOLT:HIGH

250

Query Syntax

[SOURce:]VOLTage:HIGH?

Returned Parameters

Related Commands

[SOUR:]VOLT:LEV

VOLTage:RANGe[:LEVel]

This command sets the voltage range of the power source. On the MX/RS/BPS Series, up to
three voltage ranges are available:

• 150 V AC or 200 V DC range
• 300 V AC or 400 V DC range
• 400 V AC or other (-XV) range

The LIM:VOLT? Query may be used to determine which AC voltage ranges are available to
select. This command always returns the AC voltage range values. When in DC mode,
used 200 iso 150 or 400 iso 300 to select the low and high DC voltage range. Sending a
parameter other than these will not be accepted.