Total exhaust volume, Exhaust (inflow) alarm setpoint – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 96
![background image](/manuals/738628/96/background.png)
Page 96 of 197
Total Exhaust Volume
The total volume of air exhausting from the cabinet can be determined as follows:
Note: The dress panel MUST be installed on the cabinet to obtain an accurate inflow
value.
Note: The exhaust volume noted with a DIM on the front of the cabinet may not agree
with the value obtained from a traditional Pitot tube or thermal anemometer traverse
reading of the exhaust duct.
1. Shut off the cabinet blower.
2. Remove the prefilter element from the top of the cabinet.
3. Completely seal the air intake opening with tape or plastic film secured with
tape.
4. Seal a DIM to the face of the Logic.
5. Record at least 10 inflow volumes, and average the results. The total exhaust
volume should approximately equal the value shown in Table 6-6.
Table 6-6
Total Exhaust Volume Specifications
Model 34410
34610
DIM-measured Exhaust
(CFM)
665+/-15 998+/-20
Traverse-measured
Exhaust (CFM)
810 1218
Exhaust (Inflow) Alarm Setpoint
For units with Firmware of Revision G or later, the exhaust alarm setpoint must be
entered during the calibration procedure. To do this:
1. With the cabinet operating at its nominal inflow in the calibration mode at the
alarm setpoint screen, ensure that a properly operating DIM or thermal
anemometer (for secondary inflow) is properly positioned at the face of the
Logic.
2. Close the exhaust damper until the inflow equals 80% of the total exhaust
volume, as shown in Table 6-7.
3. Press the ‘OK’ button when the proper inflow is obtained.
4. Return the damper to the position required for nominal inflow and lock the
damper in position.
Table 6-7
Inflow Alarm Volume Specifications
Model 34410
34610
Nominal Inflow (CFM)
283 +/-13
423+/-20
Inflow alarm setpoint volume (CFM)
150 +/- 5
223 +/- 5
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx