Testing a circuit breaker, Replacing a circuit breaker – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 143

Page 143 of 197
Circuit Breaker Specifications, 230 VAC
Model
34308 34300 34408 34400 34508 34500 34608 34600
Motor/Light Breaker Rating,
Amps
4 4 4 4 5 5 5 5
Motor/Light Breaker, Labconco
Part #
1327209 1327209
1327209
1327209
1327204 1327204
1327204
1327204
Motor/Light Breaker,
Mechanical Products Part #
1600-062-
040-007
1600-062-
040-007
1600-062-
040-007
1600-062-
040-007
1600-062-
050-007
1600-062-
050-007
1600-062-
050-007
1600-062-
050-007
Outlet
Breaker
Rating,
Amps
3 3 3 3 3 3 3 3
Outlet Breaker, Labconco
Part #
1327200 1327200
1327200
1327200
1327200
1327200
1327200
1327200
Outlet Breaker, Mechanical
Products Part #
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
1600-062-
030-007
Testing a circuit breaker
Note: If the breaker is tripped (white button is extended), press the reset button
of the breaker. If it does not reset smoothly, the breaker has failed and must be
replaced.
1. With the breaker reset, unplug the Logic, either at the wall, or at the power
cord inlet at the back of the electronics module, as shown in Figure 9-2.
2. Remove the two 7/16-inch acorn nuts and lock washers that secure the
module.
3. Tip the module on its side to access the breakers.
4. Using a continuity meter, check the continuity (resistance) across the
terminals of the breaker being tested.
5. The resistance across the breaker should not exceed the resistance noted
when the meter probes are connected to each other.
Replacing a circuit breaker
1. Unplug the Logic, either at the wall, or at the power cord inlet at the back of
the electronics module, as shown in Figure 9-2.
2. Remove the two 7/16-inch acorn nuts and lock washers that secure the
module.
3. Tip the module on its side to access the breakers.
4. Disconnect the wires connected to the breaker terminals.
5. Using a pair of pliers, loosen the chrome ring nut outside of the reset button.
Remove the ring nut, and pull the breaker out of the electronics module.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx