Towel catch removal – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 134
Page 134 of 197
Towel Catch Removal
Although not normally required, the towel catch can be removed for cleaning,
inspection, etc.
Note: The work surface of the cabinet and the towel catch must be thoroughly
decontaminated before removing either.
6. Remove the work surface as described above.
7. Lift the front end of the center support straight up to release it from the locating
pins (A). Set the support aside in the work area.
8. Remove the towel catch by pivoting the bottom out toward you (B).
9. Surface decontaminate the towel catch before removing it.
10. Reinstall the towel catch by sliding it back into position, ensuring that the two slots
on either end of the towel catch properly engage the rails on the rear of the duct, as
shown in Figure 8-19. Also ensure that the DOP Sampling Line (the clear tube with a
black cap) either passes through a hole in the towel catch, or passes along the
outside edge of the catch without being kinked.
Figure 8-19
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx