Changing the uv lamp, Uv lamp socket replacement – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 140

Page 140 of 197
Changing the UV Lamp
Note: For optimum performance, the UV lamp should be changed on an annual
basis.
Caution: The UV lamp and the work area of the cabinet must be thoroughly
surface decontaminated before removing the lamp.
1. Unplug the Logic.
2. Raise the sash to its full open position.
3. Surface decontaminate the UV lamp and the work area of the cabinet.
4. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and
out of its sockets.
5. Install new lamp by reversing the removal procedure.
UV Lamp Socket Replacement
Note: The UV lamp sockets are left- and right-handed. They are not
interchangeable. The Part number for the right socket assembly is
3742000, and the left is part number 3742001.
1. Unplug the Logic
2. Raise the sash to its full open position.
3. Surface decontaminate the work area and the UV lamp.
4. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and
out of its sockets.
5. To replace a socket, Use a #1 Phillips screwdriver to remove the screw.
6. Pull the socket assembly out of the rear wall.
7. Disconnect the wire connectors, and handle the old socket as contaminated
waste.
8. Install the new socket.
9. Reinstall the Phillips screw through the rear wall and capture the lamp socket
bracket. Tighten the screw.
10. Reassemble the unit.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx