Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 172

Page 172 of 197
UV Lamp Kit No. 3858500, -01, -02
All Labconco Purifier Logic Series Biosafety Cabinets are pre-wired for the installation of
a UV lamp kit, consisting of the UV lamp, its ballast, and fasteners. The kit is part
number 3858500 for 115 Volt models, 3858501 for 100 Volt models, and 3858502 for
230 Volt units.
A qualified electrician or certifier should install the UV lamp kit.
Decontaminate all interior work surfaces of the Purifier before beginning the
installation of the UV lamp.
Installation Procedure
1. Disconnect power to the unit.
2. Using a 7/16-inch wrench or socket, remove the two acorn nuts that secure the
electronics module to the top of the Purifier.
3. Tilt the electronics module on its side to access the underside of the module.
Secure the module so that it does not fall and damage the wiring.
4. Locate the fluorescent lamp ballast in the module. The UV lamp ballast mounts
next to the fluorescent ballast, as shown below.
UV Ballast
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx