Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 60

Page 60 of 197
Security Lock
NOTE: The security lock automatically reactivates and locks out the
keypad every time the blower is shut off, by the keypad or by closing the
sash. After the initial auto diagnostic screen is cleared, the security lock
reactivates.
The security lock feature prevents operation of the Biosafety Cabinet by
unauthorized users. To engage the security lock, access the Security Lock
submenu, as described in the Logic Menu Instructions, Figure 4-5. Select the
“secure” option, and press the “OK” button. A ‘lock’ icon appears on the LCD
display. Once engaged, the operator must enter the proper sequence of keystrokes
to operate the cabinet. The proper sequence to unlock the Security Lock is:
Press the “▼” keypad 3 times within 2 seconds.
After successful entry of the unlock code, the lock icon disappears from the display,
and the user has full access to the keypad functions.
To disable the Security Lock, enter the Security Lock submenu and select the non-
secure mode. The lock is completely disabled.
Filter Life Display Options
This menu selection determines how the filter life remaining gauge is displayed.
When “Bar Graph with %” is selected, the gauge will be displayed as a horizontal
bar graph with the filter life remaining shown graphically and digitally. When “Circle
Graph” is selected, the gauge is displayed as a semicircular bar graph, including a
digital display.
Data Output Rate
Note: This selection will only work if the optional RS-232 board is
installed.
This menu option selects the rate that the RS-232 board outputs data. Data can be
output at a rate of once per second, once every 10 seconds, once every 30
seconds, or once per minute.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx