PROBLEM
CAUSE
CORRECTIVE ACTION
Fluorescent light
is dim or
flickering
Lamp(s) are
defective
Replace defective lamp(s)
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective lamp
ballast
Replace lamp ballast.
UV light not
working
Sash is open
Close sash – UV light will not work
with the sash open.
Lamp is defective
Replace defective lamp.
Lamp wiring is
disconnected.
Inspect lamp wiring.
Defective
lamp
ballast
Replace lamp ballast.
Keypad
disconnected or
defective
Run keypad diagnostics and check
connections.
UV light is dim or
flickering
Lamp is defective
or is at end of
operating lifetime.
Replace defective or worn out lamp.
Lamp wiring is
disconnected
Inspect lamp wiring.
Defective lamp
ballast
Replace lamp ballast.
Airflow Alert
goes off and/or
there is a slight
decrease in filter
life remaining
gauge
HEPA filter loading
The gauge reading steadily
decreases as the Cabinet is used.
Blockage
of
the
return air slots or
grille
Check all return air slots and grilles
to ensure that they are not blocked
or restricted.
Blockage of the
Ensure that the exhaust outlet is
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx