Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 115

Page 115 of 197
Figure 7-23
ECM Power Connector Release
Tabs
ECM Data Connector Release Tab
7. Using a 1/2-inch wrench, remove the two upper and one lower blower mounting
bolts, as shown in Figure 7-24.
Figure 7-24
Blower Mounting Bolts
The motor/blower assembly is heavy. Handle with care.
8. While supporting the blower assembly, pull the assembly straight out the
plenum, as shown in Figure 7-25.
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx