4. Locate a sampling port downstream of the exhaust HEPA filter. The sampling
port should ideally be at least 10 duct diameters from any elbow and 10 duct
diameters downstream of the filter.
5. The Logic is now ready for the Exhaust HEPA Filter Leak Test. Set the
photometer sampling valve to “DOWNSTREAM”. Ensure there is a proper
vacuum at the sampling tube.
6. Open the Laskin nozzle(s).
7. Ensure the pressure gauge on the generator is reading 20 +/- 1 PSIG.
8. Observe the average concentration of aerosol downstream.
Acceptance
Average aerosol penetration shall not exceed 0.005%.
Reassembly
Reinstall the prefilter on the towel catch, if so equipped.
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx