beautypg.com

Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 163

background image

Page 163 of 197

8. Reinstall the side panel and reconnect the Logic to electrical service. The unit is

now ready for the calibration of the air flow sensor.

Note: You will need to access the certifiers menu to calibrate the airflow
sensor.

Note: To properly calibrate the air flow sensor, a qualified certifier must
verify the inflow and downflow values.

With the Logic operating, press:


“Menu”


The Display will change to Menu mode. Press the up or down arrows

(▲

or

▼)

until

“Cabinet Setup” is selected. Press:

“Mute/OK”

The Display will then change to the password mode. The password is a sequence of
keys, press:


“Light”
“UV Light”
“Timer”
“Timer”

“Mute/OK”

The display will now change into the speed setting mode.

To Change the Blower Speed


In the speed setting mode, the number shown is the motor power factor, and

should be 65-85. To increase or decrease the blower speed, press the up or

down (▲

or

▼) arrows, respectively. When the number is properly set, wait

until the display stops flashing “wait”, and press:

“Mute/OK”
This resets the blower speed, and the display will change to the filter gauge

mode.

To Bypass Changing the Blower Speed

Press:
“Mute/OK”

This leaves the blower speed as it is, and the display will change to the filter
gauge mode.