8. Reinstall the side panel and reconnect the Logic to electrical service. The unit is
now ready for the calibration of the air flow sensor.
Note: You will need to access the certifiers menu to calibrate the airflow
sensor.
Note: To properly calibrate the air flow sensor, a qualified certifier must
verify the inflow and downflow values.
With the Logic operating, press:
“Menu”
The Display will change to Menu mode. Press the up or down arrows
(▲
or
▼)
until
“Cabinet Setup” is selected. Press:
“Mute/OK”
The Display will then change to the password mode. The password is a sequence of
keys, press:
“Light”
“UV Light”
“Timer”
“Timer”
“Mute/OK”
The display will now change into the speed setting mode.
To Change the Blower Speed
In the speed setting mode, the number shown is the motor power factor, and
should be 65-85. To increase or decrease the blower speed, press the up or
down (▲
or
▼) arrows, respectively. When the number is properly set, wait
until the display stops flashing “wait”, and press:
“Mute/OK”
This resets the blower speed, and the display will change to the filter gauge
mode.
To Bypass Changing the Blower Speed
Press:
“Mute/OK”
This leaves the blower speed as it is, and the display will change to the filter
gauge mode.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx