9. Route the sensor cable along the front of the cabinet and secure it to the
existing harness running along the front of the cabinet, using nylon ties along
the top of the cabinet as shown. Ensure that there is slack in the sensor cable
where it connects to the harness.
NOTE: Ensure the sensor harness does not contact any
moving parts.
Existing Wiring Harness
One nylon tie
(more ties are shown below)
Slack in the sensor cable
10. Remove the two Phillips screws on the front edge of the right side corner post,
and allow the front of the side panel to swing out and lift off.
11. Continue routing the sensor cable
along the main harness on the right
side of the cabinet. Secure the
sensor harness with the remaining
ties as shown.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx