Purifier type a2 logic service operations, Recommended tools for service, External dress components – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 101: Front dress panel removal and installation

Page 101 of 197
Section Seven –
Purifier Type A2 Logic Service Operations
This section will review the tools needed, and common service operations for the Logic.
Recommended Tools for Service
The tools needed for most common service operations are:
9
Logic Certifier Kit, Labconco Part # 3858400
9
#2 Phillips Screwdriver or long bit for cordless drill
9
½-inch Socket with extension and ratchet
9
½-inch nut driver
9
7/16-inch socket or nut driver
9
5/16-inch socket or nut driver
Optional tools that may be needed include:
9
Cordless screwdriver/drill
9
Cordless ratchet
9
3/8-inch nut driver or socket and ratchet
9
Pliers
9
10-24 Tap
9
5/16 x 18 Tap
Note: If you use a cordless drill with an adjustable torque setting, set the
torque as low as possible to properly drive the screw or bolt.
External Dress Components
Front Dress Panel Removal and Installation
Figure 7-1
1. Locate and remove the two Phillips screws that
secure the front panel as shown in Figure 7-1.
They are located on the bottom corners of the
front
dress
panel
2.
Swing the bottom of the dress panel out to
clear the fluorescent light and then lift the
front dress panel straight up and away from
the
cabinet.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx