Uv lamp timer, Uv lamp maintenance – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 57
![background image](/manuals/738628/57/background.png)
Page 57 of 197
UV Lamp Timer
The UV Lamp Timer allows you to decide whether the UV lamp will operate continuously
or for a time period that you select.
Interval
When the “Interval” option is selected, the UV Interval Timer menu is displayed.
UV Interval Timer
This window lets you set the amount of time the UV lamp to stays lit
when activated. Use the “▲” and “▼” buttons to select the amount of
time you need, and then press the “OK” button to set the interval.
NOTE: Raising the sash will reset the timer to its original setting.
Pressing “UV Light” will turn the light off, pressing
“Mute/OK“will pause the timer.
Continuous
In the continuous mode, the UV light will remain lit until the “UV Light” button is
pressed, or the sash is raised.
UV Lamp Maintenance
This selection allows you to check how many hours the UV lamp has burned. It also
allows you to set the number of hours you want the lamp to last, and to reset the UV
lamp hourmeter.
UV Lamp Hourmeter
This display only shows how many hours the UV lamp has been lit, and how many
hours remain until you will receive a warning to replace the lamp.
Change UV Lamp Life
In this option, you can set the number of operating hours before receiving the
replace UV lamp warning. For most UV lamps, the output of UV light decreases at a
constant rate. Typically, after 6,000 hours of operation the lamp will output 80% of
the UV light it did when it was new. This option allows you to set the warning at a
life you prefer, in 100 hour increments.
Reset UV Lamp Clock
This option lets you reset the UV hourmeter to 0 whenever the lamp has been
replaced.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx