5. Using the Phillips screws and locking nuts provided, secure the UV ballast to the
module.
6. Wire the UV ballast into the wiring harness, following the wiring diagrams located
in Figure 2a-c.
7. Reattach the electronics module to the top of the cabinet before reconnecting
the electrical power.
8. Raise the sash, and install the UV lamp by rotating the lamp pins into the
sockets.
9. Lower the sash to fully closed position, plug in the unit, and turn the system
power switch to the "UV" position. The UV lamp will illuminate. Refer to the
user’s manual for further information regarding the proper operation and
maintenance of the UV light.
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx