Acceptance
Aerosol penetration shall not exceed 0.01%.
Reassemble the unit by compressing the exhaust filter cover while installing its tabs
back into the exhaust filter bracket slots. Reinstall the diffuser, ensuring that the holes in
the back of the diffuser frame engage the posts on the rear wall of the work area, as
shown in Figure 5-6.
Figure 5-6
Diffuser pins engaging the holes in the
rear of the diffuser.
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx