Installing the purifier on a labconco base stand – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 40

Page 40 of 197
Figure 3-2c
Installing the Purifier on an Existing Work Surface
Note: The Purifier is very top heavy. Use caution when lifting or moving it.
Note: The Purifiers weigh from 400 to 700 pounds. Always use safe lifting
practices when picking it up.
When installing the Purifier onto an existing work surface or benchtop, ensure that the
structure can safely support the combined weight of the cabinet and any related
equipment. The work surface should be at least as wide as the unit and 31 inches deep
to properly support the unit.
The work surface should be smooth and durable. The surface should be nonporous and
resistant to the disinfectants and chemicals used in conjunction with the Purifier. This
will allow a proper seal to form between the bottom of the Purifier and the work surface.
A hole or notch may need to be cut in some work surfaces in the right front corner to
accommodate the drain valve.
Installing the Purifier on a Labconco Base Stand
Labconco offers a variety of accessory base stands in a number of configurations to suit
the user’s particular needs. Stands can be ordered with preset telescoping legs or, with
a manually or electrically adjustable hydraulic lift stand.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx