Sash Cable Retainer
The sash cable retainers are located at the top of each sash pulley, as shown in
Figure 8-9. If the retainer needs adjustment, Use a 3/8-inch wrench to loosen
the bolt securing the retainer and adjust it until it contacts the pulley. Retighten
the bolt.
Figure 8-9
Exhaust Damper Adjustment
The exhaust damper adjustment blade is located on exhaust stack. Using a #2 Phillips
screwdriver, loosen the damper locking screw to move the damper open or closed, as
shown in Figure 8-10.
Figure 8-10
Damper Locking Screw
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx