beautypg.com

Blower operation – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 59

background image

Page 59 of 197

Light Hourmeter

This display only shows the hours the fluorescent lamps have been lit and how
many hours remain until you receive a warning to replace the lamps.

Change Light Life

In this option, you can set the hourmeter before you get the replace light warning.
For most fluorescent lamps, the output of light decreases at a constant rate.
Typically, after 6,000 hours of operation the lamp will output less light than it did
when it was new. This option allows you to set the warning at what ever level you
prefer, in 100 hour increments.

Reset Light Clock

This option resets the light hourmeter to 0 after the fluorescent lamps have been
replaced.

Blower Operation

This selection allows you to choose to operate the blower in Smart-Start, Manual
Touchpad, or in Night-Smart modes.

Smart-Start

When the “Smart-Start” option is selected, the blower will turn on every time the
sash is opened. When the sash is closed, the blower will shut off.

Manual Touchpad

When the “Manual Touchpad” option is selected, the blower must be turned on and
off at the touchpad.

Night-Smart

When the “Night-Smart” option is selected, the blower will turn on every time the
sash is opened. When the sash is closed, the blower speed will decrease, to deliver
20-30% of the volume of air during normal operation.

Cabinet Set Up

These selections allow for the customization of the display and the cabinet operation.

Units of Measure

When equipped with the optional airflow sensor, this option allows you to select the
air velocity units of measure. When “Metric” is selected, the air velocities will be
displayed in meters-per-second. When “US” is selected, the velocities will be
displayed in feet per second.