Drain Valve Installation (optional)


In order to prevent damage during shipping, the drain valve assembly has not been
installed in the unit. If desired, the valve should be installed after the unit is in its final
location.
To install the valve assembly, follow these steps:
Note: The work surface is heavy. Use caution when handling it.

1.

Lift the work surface out of the Purifier by lifting on the knobs at the front of the
work surface. Steady the work surface while pulling it straight out of the cabinet.

2.

Using a putty knife, remove and discard the stainless steel cover that is sealed over
the drain mounting holes. Scrape out remaining sealant that is around the holes.

3.

Apply a light coating of the silicone sealant (user supplied) to the mounting surface
of the drain assembly. Attach the drain assembly under the bottom of the cabinet as
shown in Figure 3-6. Wipe off any excess sealant from the cabinet bottom. Ensure
that the center drain hole is unobstructed

.

4. Make sure the drain valve is in the closed position.
5. Reinstall the work surface.
6. Allow the silicone sealant to cure for at least eight hours before exposing it to liquid.

Figure 3-6

NOTE: The drain valve assembly
attaches to the underside of the cabinet
bottom.

Apply a light coat of silicone sealant to
this surface of the connector, aligning
the three holes in the connector with the
three holes in the Purifier liner.