Adjusting the inflow velocity – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 84

Page 84 of 197
Logic Filter Life Gauge Operation and Resetting for Local
Conditions
The filter life gauge calculates the remaining filter life by measuring and comparing the current
motor speed to the initial speed noted at the factory during assembly, and the calculated
maximum motor speed, using the formula:
100-(100x (current speed-original speed) ) = % filter life left
max. speed-original speed
So
anything
that changes the motor’s current speed, like:
9
Blocking the grille.
9
Raising the air temperature.
9
Lowering the barometric pressure.
will increase the motor’s speed, causing the Filter Life Gauge to decrease. Electronic dampening
of the gauge will limit the change in readings to approximately 1%/minute.
Logic Biosafety Cabinets being installed at altitudes or operating at temperatures significantly
different from those at the time of manufacture (900 +/- 500 ft above sea level, 75
o
+/- 5
o
F)
must
have their filter life gauge recalibrated for local conditions. Failure to do so will result in the
users noticing a steady drop in Percent (%) Filter Life remaining as the unit operates. Below are
the instructions on how to correct the Filter Life Gauge for local conditions.
8. Ensure to correct all measurement equipment for local conditions (temperature, absolute
pressure). Anemometers may not have this ability and might have to be done manually;
whereas, a ShortRidge
TM
instrument has a setting to correct for local conditions.
9. As the elevation or the ambient temperature rises, the motor power factor will have to be
reduced to compensate. For an initial certification, the motor speed should be reduced
until it equals the Logic’s “Initial Motor Speed” as reported on the unit’s test sheet, plus
approximately 10-15 RPM per each additional 1000 feet of elevation.
10. The Logic MUST be recertified using instruments corrected for local conditions.
11. When the unit is recertified, and operating properly, the certifier MUST RESET THE
FILTER LIFE GAUGE, INDICATING A NEW FILTER HAS BEEN INSTALLED.
12.
Only by performing Step #4 AFTER THE UNIT IS READJUSTED AND RECERTIFIED FOR
ITS LOCAL CONDITION, will the filter life gauge register 100% (+/- 2%) during
operation. As the Filters then load, the filter life gauge will then begin to drop as
appropriate.
13. If you do not wish to reset the filter gauge, select the option “Filter not changed”, and
press “Mute/OK”.
14. Either selection will lead to calibration of the inflow velocity.
Adjusting the inflow velocity
1. Using the “▲” or “▼” buttons, increase or decrease the displayed inflow velocity
until it matches the observed value, then press “Mute/OK” to confirm the value.
2. Once the downflow is confirmed, the unit will then automatically turn off and
then back on.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx