Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 192

Page 192 of 197
exhaust outlet
not blocked or restricted.
Blockage
or
restriction under
the work surface
Ensure that the towel catch and
plenum beneath the work surface
are unobstructed.
Contamination of
work in the
cabinet
Improper technique
or procedure for
the Biosafety
Cabinet
See “Use of the Cabinet” section in
the manual.
Restriction
of
the
return air slots or
grille – blockage of
the exhaust outlet
Ensure that all return air slots,
grilles and the exhaust outlet are
unobstructed.
External factors are
disrupting the
cabinet airflow
patterns or acting
as a source of
contamination
See “Installation” section of this
manual.
Cabinet is out of
adjustment/HEPA
filter(s) are
defective
Have cabinet recertified.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx