6. Using a 1/2 -inch socket or wrench, loosen the four exhaust filter sealing bolts
located at each corner of the exhaust filter tray, as shown in Figure 7-17, at least
1/2 inch. This releases the exhaust HEPA filter from the cabinet.
Figure 7-17
Exhaust HEPA Filter bolts
(2 per each side)
7. Pull the exhaust HEPA filter straight out of the cabinet.
To Remove the Exhaust HEPA Filter on 3-foot models:
1. Follow steps 1-5 as above.
2. Loosen and remove the four sealed exhaust filter bolts on the top of the cabinet
hull, as shown in Figure 7-18 and 7-19.
3. Remove the exhaust filter pan by lifting it straight up and then out of the
cabinet, as shown in Figure 7-20.
4. When reinstalling the pan, ensure all four edges of the pan are OUTSIDE the
sealing edges of the plenum, as shown in Figure 7-21.
Figure 7-18
Exhaust HEPA Filter bolts
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx