On Models Manufactured After April 2009 –
1. Turn off the Biosafety Cabinet.
2. Remove the exhaust HEPA filter cover by loosening the four Phillips
Screws on the sides of the cover, then pivot the front of the cover up, and
then pull it forward as shown.
Pivot the front of the cover up and then pull it
forward (dress panel removed for clarity only).
Loosen these screws only
Note: exercise caution when removing or installing the exhaust filter cover, to
prevent damaging the Exhaust HEPA filter.
3. Locate and remove the four acorn nuts
and lockwashers that secure each exhaust
cover bracket to the top of the Logic.
Keep the nuts and washers; you may
discard the brackets.
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx