Quickchart - logic type a2 biohazard cabinets – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 16

Page 16 of 197
QuickChart - Logic Type A2 Biohazard Cabinets
Model
34308 34300 34408 34400 34508 34500 34608 34600
Type
A2 A2 A2 A2 A2 A2 A2 A2
Cabinet
Size
(in
feet) 3 3 4 4 5 5 6 6
Sash
Opening
(inches) 8 10 8 10 8 10 8 10
Starting Serial #
0705_
0705_
0705_
0705_
0705_
0705_
0705_
0705_
HEPA Filter Leak Test Data
Air
Displacement
(CFM) 543 596 723 794 903 991 1083 1189
Laskin
Nozzles
needed 1 1 2 2 2 2 2 2
Theoretical aerosol conc. (ug/l)
1
25 23 38 34 30 27 25 23
Actual aerosol conc. (ug/l)
2
14 12 21 18 17 15 17 15
Nominal
Speed
Control
Setting 50-60 55-65 65-75 75-85 75-85 80-90 65-75 75-85
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5 55+/-5
55+/-5 55+/-5
55+/-5
Downflow
test
grid
#
of
points
21 21 24 24 30 30 36 36
Test Grid (# rows x # columns)
3 x 7
3 x 7
3 x 8
3 x 8
3 x 10
3 x 10
3 x 12
3 x 12
Test Grid Data (in inches)
Grid distance from back & sides
6.0 6.0
6.0
6.0
6.0
6.0
6.0
6.0
Distance between rows
5.75
5.55 5.75 5.55 5.75 5.55 5.75 5.55
Distance between columns
4.08 4.08 5.21 5.21 5.39 5.39 5.50 5.50
Inflow Data
Nominal
Average
Inflow
(FPM) 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5 105+/-5
Sash Open Area (Sq. Ft)
2.03 2.53 2.69 3.37 3.36 4.2 4.03 5.03
Nominal Avg. Exhaust Vol. (CFM)
213 266 283 354 353 441 423 529
Avg.
Exhaust
Vol.
Range
(CFM) 203-223 253-278 269-296 337-371 336-370 420-462 403-443 503-553
Secondary Inflow Data
3
Restricted Sash Opening (inches)
4
4 4 3 3
2.25
2.25
4.25
4.25
Sensor distance (inches)
5
4
4 3.25 3.25 3.25 3.25 4
4
# of Test points
6
6
8
8
10
10
12
12
Test point location
*
*
*
*
*
*
*
*
Nominal
Avg.
Inflow
Vel.
(FPM)
213 266 283 354 353 441 231 289
Avg.
Inflow
Vel.
Range
(FPM) 203-223 253-278 269-296 337-371 336-370 420-462 220-242 275-302
Correction
Factor
(CF) 1 1 1 1
0.96
1
1.83
1.83
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
203-223 253-278 269-296 337-371 336-370 420-462 403-443 503-553
Sash Open Area (Sq. Ft)
2.03 2.53 2.69 3.37 3.36 4.2 4.03 5.03
Inflow Velocity Range
(AIV / Sash open area)
100-110 100-110 100-110
100-110 100-110
100-110 100-110
100-110
Supply HEPA Data
Width
(inches)
36 36 48 48 60 60 72 72
Depth (inches)
18
18
18
18
18
18
18
18
Height (inches)
6
3.06
3.06
3.06
3.06
3.06
3.06
3.06
3.06
Performance
(CFM)
335 335 445 445 555 555 665 665
Performance (Pressure in. H
2
O) .55-.69 .55-.69 .43-.57 .43-.57 .43-.57 .43-.57 .43-.57 .43-.57
Labconco
P/N
3838400 3838400 3838401 3838401 3838402 3838402 3838403 3838403
Filtration Group P/N
42665L
42665L 42546L 42546L 42663L 42663L 42666L 42666L
AAF P/N
3061389
-001
3061389
-001
3061389
-002
3061389
-002
3061389
-003
3061389
-003
3061389
-005
3061389
-005
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx