Optional ultraviolet lamp, Lamp socket replacement, Lamp identification – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 139: Uv lamp specifications
Page 139 of 197
Lamp Socket Replacement
1. Unplug the Logic
2. Remove the dress panel as noted in Figure 7-1
3. Remove the fluorescent lamps from the sockets.
4. To replace a socket, insert a small straight bladed screwdriver into the socket
near the lamp wire, as shown in Figure 9-1.
5. Press the wire into the new socket.
6. Reassemble the unit.
Figure 9-1
Optional Ultraviolet Lamp
Lamp Identification
All Logics use a single 30-Watt germicidal lamp. The lamp information is shown
below. The code of the lamp part number is as follows:
G30T8
G = germicidal lamp
30 = the lamp Wattage
T8 = the lamp mounting configuration
UV lamp Specifications
• For all models, the UV lamp number is G30T8. The Labconco part number
is 1271300.
• For all 115/230 VAC models, the ballast assembly is Labconco part
number 3829901. The ballast is Robertson part number
PSM2GPH18MVW.
• For all 100 VAC models, the ballast assembly is Labconco part number
3830600. The ballast is Robertson part number RSO1GPH30100.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx