Quickchart - logic type a2 biohazard cabinets – Labconco PURICARE Procedure Station Models 34812xx User Manual
Page 18

Page 18 of 197
QuickChart - Logic Type A2 Biohazard Cabinets
footnotes
Logic Model Identification
9
The primary serial tag is on the lower outside edge of the right corner post.
9
The secondary serial tag is located on the front of the electronics module on the top right
side of the cabinet.
9
The first two digits of the serial number are the year of production; the next two are the
month. The next 5 digits are the sequence of production, and the letter following the
serial number is the revision level of the cabinet.
HEPA Filter Leak Test Specifications
1. Based on mineral oil.
2. The actual aerosol concentration is what was observed during testing.
9
Aerosol generator should be placed in the left rear corner of the work area, pointing
toward the rear grille. For uncontaminated units, the upstream concentration can be
sampled from the tube located under the work surface.
Downflow Test Specifications
9
All models are classified as uniform downflow.
9
All tests performed as described in ANSI/NSF Standard 49:2007.
9
UV Lamp, IV bar and all other accessories must be removed before measuring downflow.
Inflow Test Specifications
9
All tests performed as described in ANSI/NSF Standard 49:2007.
Secondary Inflow Test Specifications
3. Must use Labconco holder P/N 3836405 to perform this test properly.
4. Measured from the bottom edge of the sash to the mating surface of the grey corner
post end cap.
5. Measured from the bottom edge of the probe holder to the center of the thermal
anemometer sensor element.
* - Locate the single row of holes at the front of the grille. Mark the 6
th
hole from the side
wall and subsequent test points every 9 holes until the number of test points marked equals
the width of the cabinet in feet (3-foot cabinet, mark the first 3 points). Repeat for the
opposite side.
Supply and Exhaust HEPA Filter Specifications
6. Without gasket
Motor / Blower Specifications
7. Each motor must be programmed by Labconco for the appropriate width cabinet. The
speed control settings will fluctuate depending on local temperature and pressure.
UV lamp Specifications
9
For all models, the UV lamp number is G30T8. The Labconco part number is 1271300.
9
For all 115/230 VAC models, the ballast assembly is Labconco part number 3829901. The
ballast is Robertson part number PSM2GPH18MVW.
9
For all 100 VAC models, the ballast assembly is Labconco part number 3830600. The
ballast is Robertson part number RSO1GPH30100.
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx