ADJUSTING THE BLOWER SPEED & FILTER GAUGE
Type A2 without airflow sensor
Press the or keys to adjust the blower
speed. The word ‘wait’ will flash until the
blower speed stabilizes. Press ‘OK’ to
accept the new setting.
CERTIFIER USE ONLY
Calibration mode:
Motor Speed Factor:
wait
900
950
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST
80
CERTIFIER USE ONLY
Calibration mode:
Filter setup:
1. New Filter
2. Filter Not Changed
UP/DOWN TO CHANGE, OK
TO NEXT, MENU TO LAST
Press the or keys to select. Option # 1 will
reset the filter life remaining gauge to 100%.
Option # 2 will leave the gauge unchanged.
Press ‘OK’ to accept your choice.
The Logic will automatically
turn off and reset itself into
normal operation.
This manual is related to the following products:
- PURICARE Procedure Station Models 34810xx PURICARE Procedure Station Models 34802xx PURICARE Procedure Station Models 34800xx Type B2 Biosafety Cabinets Models 34610xx Type B2 Biosafety Cabinets Models 34410xx Type A2 Biosafety Cabinets Models 34609 Type A2 Biosafety Cabinets Models 34509 Type A2 Biosafety Cabinets Models 34409 Type A2 Biosafety Cabinets Models 34309 Type A2 Biosafety Cabinets Models 34608xx Type A2 Biosafety Cabinets Models 34508xx Type A2 Biosafety Cabinets Models 34408xx Type A2 Biosafety Cabinets Models 34308xx Type A2 Biosafety Cabinets Models 34600xx Type A2 Biosafety Cabinets Models 34500xx Type A2 Biosafety Cabinets Models 34400xx Type A2 Biosafety Cabinets Models 34300xx