Section RX (SONET)
Alarm Analysis
³
SEF (Severely Errored Framing): A SEF defect indicates that a
minimum of four consecutive errored framing patterns are received.
³
LOF (Loss Of Frame): A Loss Of Frame alarm indicates that a Severely
Error Framing (SEF) defect on the incoming SONET signal persists for
at least 3 milliseconds.
³
TIM-S (Trace Identifier Mismatch - Section): The TIM-S defect indicates
that the received J0 Trace doesn’t match the expected message value.
The TIM-S alarm is only available when Enable TIM-S check box from
J0 Trace section has been selected.
J0 Trace
³
Received Message: Displays the received J0 value. The R > and F >, represent respectively a carriage return and a line feed. ³ Enable TIM-S (Trace Identifier Mismatch - Section): Allows enabling ³ Expected Message: Allows entering the expected J0 Trace message. J0 ³ Expected Format: Allows the selection of the expected format: 16 or
represents the CRC-7 for a 16-bytes format. The last two bytes of a
64-bytes format,
the Trace Identifier Mismatch for the expected message defined.
Enable TIM-S has to be enabled to give access to the expected trace
format and message.
value should be ASCII suitable characters. The default message is
EXFO SONET/SDH for 16 bytes and EXFO SONET/SDH Analyzer
Section/RS trace test message for 64 bytes.
64 bytes. The default setting is 16 bytes.