Temperature characteristics, Slow rate, Display in °c, °f, exclusive of probe errors.) – GW Instek GDM-8261A User manual User Manual
Page 225
APPENDIX
225
[ 6 ] Need Fixed ACI/ACV range and Trigger Delay=0.
[ 7 ] To meet the specifications accuracy,
Slow rate
setting is needed.
Temperature Characteristics
(Display in °C, °F, Exclusive of probe errors.)
RTD
[1]
(Accuracy based on PT100):
(100Ω platinum [PT100], D100, F100,PT385, PT3916, or user type)
Range
Resolution
1 Year (23°C ±5°C)
Temperature Coefficient
0°-18°C & 28°-55°C
-200°C~ -100°C
0.001°C
0.09°C
0.004 °C / °C
-100°C~-20°C
0.001°C
0.08°C
0.005 °C / °C
-20°C~20°C
0.001°C
0.06°C
0.005 °C / °C
20°C~100 °C
0.001°C
0.08°C
0.005 °C / °C
100°C~300 °C
0.001°C
0.12°C
0.007 °C / °C
300°C~600 °C
0.001°C
0.22°C
0.009 °C / °C
Thermocouples
[1]
(Accuracy based on ITS-90):
Type Range
Resolution
90 Day/1 Year
(23°C±5°C)*
Temperature Coefficient
0°-18°C & 28°-55°C
E
-200 to +1000°C
0.002 °C
0.2 °C
0.03 °C / °C
J
-210 to +1200˚C
0.002 °C
0.2 °C
0.03 °C / °C
T
-200 to +400°C
0.002 °C
0.3 °C
0.04 °C / °C
K
-200 to +1372˚C
0.002 °C
0.3 °C
0.04 °C / °C
N
-200 to +1300˚C
0.003 °C
0.4 °C
0.05 °C / °C
R
-50 to +1768˚C
0.01 °C
1 °C
0.14 °C / °C
S
-50 to +1768˚C
0.01 °C
1 °C
0.14 °C / °C
B
+350 to +1820˚C
0.01 °C
1 °C
0.14 °C / °C
*Relative to simulated junction
Reading Rate
[2]
(Readings/sec) TCO/ RTD
Rate
Digits
Rate
Slow
6 ½
10
Mid
5 ½
60
Fast
4 ¼
300
[1] Specifications do not include probe accuracy
[2] All speeds need A-Zero=off, A-Gain=off, Fixed range and Trigger Delay=0.