REMOTE CONTROL

205

ROUTe Commands

ROUTe:CLOSe

Close a specified scanner channel.
Parameter: (101~118)
Example: ROUT:CLOS 102
Closes channel 102.

ROUTe:OPEN:ALL

Opens all scanner channels.

ROUTe:MULTiple:OPEN

Enable all channels in a specified range. Channels that are not in the range
are not affected.
Parameter: (101~118)
Example: ROUT:MULT:OPEN 105,110
Channels 105 to 110 are enabled.

ROUTe:MULTiple:STATe?

Returns the status of all the scanner channels that are open.
Return parameter: 101 OFF, 102 ON, 103 ON etc.

ROUTe:MULTiple:CLOSe

Disable channels in a specified range.
Parameter: (101~118)
Example: ROUT:MULT:CLOS 105,110
Disables channels 105~110.

ROUTe:FUNCtion

Enables scan related functions
Parameter: OFF | SCAN | STEP
Example: ROUT:FUNC SCAN
Enables the SCAN function.

ROUTe:FUNCtion?

Returns the Scan related function status.