beautypg.com

GW Instek GDM-8261A User manual User Manual

Page 108

background image

GDM-8261A User Manual

108

2. Route the wires as shown below via the two openings

(left and right) at the front cover.

Note

When using thermocouple wiring, please use

extension wires so that the cold junction points are
external to the scanner card. Connecting

thermocouple wiring directly to the scanner box is
not recommended due to the radiant heat from the

internal components.

Best if longer

than 20cm

Thermocouple

wire

Cold junction

points

Extension

wires

Hot junction

point

3. Bundle the wires at the front cover using the holes at

the bottom.