REMOTE CONTROL

175

CONFigure2:VOLTage:AC

Sets measurement to AC Voltage on the second display and specifies
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF2:VOLT:AC
Sets the measurement mode to AC voltage.

CONFigure2:CURRent:DC

Sets measurement to DC Current on the second display and specifies
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF2:CURR:DC 10e-3,DEF
Sets the DC current range to 10mA using the default resolution on the
second display.

CONFigure2:CURRent:AC

Sets measurement to AC Current on the second display and specifies
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF2:CURR:AC 10e-2,MAX
Sets the measurement mode to ACI with a 100mA range at the maximum
resolution.

CONFigure2:RESistance

Sets measurement to 2W Resistance on the second display and specifies
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example:

CONF2:RES 10e3,MIN

Sets the range to 10kΩ

with the lowest resolution.

CONFigure2:FRESistance

Sets measurement to 4W Resistance on the second display and specifies
the range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF2:FRES 10e3,MAX
Sets the measurement mode to 4W with a range of 10k

at the maximum

resolution.