REMOTE CONTROL

163

Command Syntax

The commands are partially compatible with IEEE488.2 (1992) and SCPI

(1994) standard. Commands are NON-case sensitive.

Example
command

C O N F

:

V O L T

:

D C

,

M A X

1

2 3 4

5

1 0

1: Command Header

2: Single space

3: Parameter 1

4: Comma (no space after
comma)

5: Parameter 2

Parameter
example

Boolean

Boolean logic: 0 or 1. Used for On (1) or

Off (0) command.

NR1

Integer: 0, 1, 2, 3.....

NR2

Decimal number: 0.0, 0.1, 0.2,....

NR3

Floating point number: 4.5e-1, 8.5e+1,...

NRf

Any NR1,NR2 or NR3 value.

MIN, MAX

The GDM-8261A automatically translates
to the Minimum (min) or Maximum (max)

value available.

DEF

Default setting value.

Automatic
parameter range
selection

The GDM-8261A automatically translates the command
parameter into the closest available value.

Example 1

CONF:VOLT:DC 1 (Sets the

measurement item to DC Voltage and the
range to 1V).

Example 2

CONF:VOLT:DC 2 (Sets the

measurement item to DC Voltage and the
range to 2V). There is no 2V range so the

GDM-8261A selects the closest range,
10V.

EOL

Marks the end of a command line. The following

messages are in accordance with IEEE488.2 standard.

LF, CR,CR+LF

EOL, user configurable (excluding GPIB),
see page 134.