REMOTE CONTROL

173

CONFigure:FRESistance

Sets measurement to 4W Resistance on the first display and specifies the
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF:FRES 10e3,MAX
Sets the measurement mode to 4W with a range of 10k

at the maximum

resolution.

CONFigure:FREQuency

Sets measurement to Frequency on the first display and specifies
range/resolution.
Parameter:

[None] | [Range( | MIN | MAX | DEF),Resolution( |

MIN | MAX | DEF)]
Example: CONF:FREQ MAX,MAX
Sets the frequency measurement range to max and the resolution to max.

CONFigure:PERiod

Sets measurement to Period on the first display and specifies the
range/resolution.
Parameter: [None] | [Range( | MIN | MAX | DEF),Resolution( |
MIN | MAX | DEF)]
Example: CONF:PER
Sets the DMM to period measurement using the previous
range/resolution.

CONFigure:CONTinuity

Sets measurement to Continuity on the first display.
Parameter: None

CONFigure:DIODe

Sets measurement to Diode on the first display.
Parameter: None

CONFigure:TEMPerature:TCOuple

Sets measurement to Temperature thermocouple (T-CUP) on the first
display.
Parameter: [None] | [Type(B | E | J | K | N | R | S | T)]
Example: CONF:TEMP:TCO J
Sets the measurement mode to TCO with a type J sensor.

CONFigure:TEMPerature:FRTD

Sets the measurement mode to 4W RTD measurement mode on the first
display. Sets the sensor type.
Parameter: [None] | [Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)]
Example: CONF:TEMP:FRTD PT100
Sets the sensor type to PT100 and sets the measurement mode to 4W RTD