REMOTE CONTROL

187

[SENSe:]VOLTage:DC:RANGe?

Returns the DC Voltage measurement range.
Parameter: [None] | [MIN | MAX]

[SENSe:]VOLTage:AC:RANGe

Sets the AC Voltage measurement range.
Parameter: ( | MIN | MAX)
Example: SENS:VOLT:AC:RANG MIN
Set the AC voltage range to lowest range allowed.

[SENSe:]VOLTage:AC:RANGe?

Returns the AC Voltage measurement range.
Parameter: [None] | [MIN | MAX]

[SENSe:]CURRent:DC:RANGe

Sets the DC Current measurement range.
Parameter: Range( | MIN | MAX)
Example: SENS:CURR:DC:RANG 10 e-2
Sets the DC current range to 100mA.

[SENSe:]CURRent:DC:RANGe?

Returns the DC Current measurement range.
Parameter: [None] | [MIN | MAX]

[SENSe:]CURRent:AC:RANGe

Sets the AC Current measurement range.
Parameter: Range( | MIN | MAX)
Example: SENS:CURR:AC:RANG 10 e-2
Sets the AC current range to 100mA.

[SENSe:]CURRent:AC:RANGe?

Returns the AC Current measurement range.
Parameter: [None] | [MIN | MAX]

[SENSe:]RESistance:RANGe

Sets the 2W resistance measurement range.
Parameter: Range( | MIN | MAX)
Example: SENS:RES:RANG 1000
Sets the resistance range to 1kΩ.

[SENSe:]RESistance:RANGe?

Returns the 2W resistance measurement range.
Parameter: [None] | [MIN | MAX]