198

VAL1?

Returns the 1

st

display reading in the unit format specified in the

Configuration menu (Return Format, page 136) or from the
SYSTem:OUTPut:FORMat command (page 200).
Example: SAMP:COUN 100
VAL1?

>+0.333E-4,V DC
>+0.389E-4,V DC
> etc, for 100 counts.

Queries 100 counts of stored samples from the 1

st

display.

VAL2?

Returns the 2

nd

display reading in the unit format specified in the

Configuration menu (Return Format, page 136) or from the
SYSTem:OUTPut:FORMat command (page 200).
Example: SAMP:COUN 100
VAL2?

>+0.345E-4,V DC
>+0.391E-4,V DC
> etc, for 100 counts.

Queries 100 counts of stored samples from the 2

nd

display.

TRIGger:SOURce

Selects the trigger source.
Parameter: INT | EXT
Example: TRIG:SOUR INT
Sets the trigger source as internal.

TRIGger:SOURce?

Returns current trigger source.

TRIGger:DELay

Sets the trigger delay in milliseconds
Parameter: (0 ~ 9999) | MIN | MAX
Example: TRIG:DEL MAX
Sets the trigger delay to the maximum.

TRIGger:DELay?

Returns the trigger delay time in milliseconds.
Parameter: None | MIN | MAX

TRIGger:AUTO

Turns Trigger Auto mode on/off.
Parameters: ON | OFF
Example: TRIG:AUTO OFF
Turns the Trigger Auto mode off.