GDM-8261A User Manual

206

ROUTe:CHANnel

Advanced configuration mode for the scanner channels. The channel
function, voltage and Auto-range mode can be configured.
Parameter: Channel(), Function(String), Range(),

Auto Range(ON|OFF)

Function: 1(VOLT), 2(VOLT:AC), 3(CURR [DCI]), 4(CURR:AC [ACI]), 7(RES),
8(FREQ), 9(TEMP:TCO:C), 13(CONT), 14(PER), 15(TEMP:TCO:F),
16(FRES), 17(DIOD), 18(TEMP:RTD:C), 19(TEMP:FRTD:C),
20(TEMP:RTD:F), 21(TEMP:FRTD:F)
Range:

Autorange: 0=Off, 1=On
Example: ROUT:CHAN 101,1,1,0
Sets channel 1 (101) to VOLT (1), 1V range (1) and disables Auto-range (0).

ROUTe:CHANnel?

Returns the advanced channel configuration settings of each channel. See
the ROUTe:CHANnel command for return parameters.
Return parameter: Channel, Function, Range, Auto Range
Example: ROUT:CHAN? 101
>

101,VOLT,0.1,ON

Returns channel 101, function is VOLT with range at 0.1V and Auto range
on.

ROUTe:COUNt

Set the number of counts for the scan.
Parameter: (1 ~ 999) | MIN | MAX
Example: ROUT:COUN 50
Sets the scan count to 50 counts.

ROUTe:COUNt?

Returns the number of counts for the scan.
Parameter: None | MIN | MAX

ROUTe:DELay

Set the Delay timer for the scan in milliseconds.
Parameter: (0 ~ 9999) | MIN | MAX
Example: ROUT:DEL 100
Sets the delay time to 100 milliseconds.

ROUTe:DELay?

Returns the Delay timer settings.
Parameter: None | MIN | MAX

ROUTe:STATe?

Queries whether the scanner box is installed or not.
Return parameter: Boolean(0|1) 0=not installed, 1=installed