GDM-8261A User Manual

196

CALCulate:STORe:COUNt

Set the number of measurement counts that are recorded with the Store
measurement function.
Parameter: (2 ~ 9999) | MIN | MAX
Example: CALC:STOR:COUN 1000
Sets the number of counts to be recorded as 1000.

CALCulate:STORe:COUNt?

Returns the number of counts that are recorded with the Store
measurement function.
Parameter: [None] | MIN | MAX

CALCulate:AVERage:COUNt

Sets the total number of statistic counts.
Parameter: (0, 2~100000) 0=continuous count, 2~100000=count
Example: CALC:AVER:COUN 0
Sets the count to continuous.

CALCulate:AVERage:COUNt?

Returns the total number of recorded counts. The setting commands for
this query are: CALCulate:STORe:COUNt, ROUTe:COUNt and
CALCulate:AVERage:COUNt.
Parameter: None | (0~2) 0=Store, 1=Scan, 2=Stats
Example: CALC:AVER:COUN? 0

>+0010

Returns the total number of counts set for the Store function (10 counts).

CALCulate:AVERage:MINimum?

Returns the minimum recorded value.
Parameter: None | (0~2) 0=Store, 1=Scan, 2=Stats

CALCulate:AVERage:MAXimum?

Returns the maximum recorded value.
Parameter: None | (0~2) 0=Store, 1=Scan, 2=Stats

CALCulate:AVERage:AVERage?

Returns the average recorded value.
Parameter: None | (0~2) 0=Store, 1=Scan, 2=Stats

CALCulate:AVERage:PTPeak?

Returns the recorded peak to peak value (max value – min value).
Parameter: None | (0|1|2) 0=Store, 1=Scan, 2=Stats
Return Parameter:

CALCulate:AVERage:SDEViation?

Returns the recorded Standard Deviation.
Parameter: None | (0~2) 0=Store, 1=Scan, 2=Stats