Unit #2 unit #1 load input input output, Output, Output com – GW Instek PSU-Series User Manual User Manual

Page 82

background image

PSU Series User Manual

82

Series Connection
to Output
Positive and
Negative Polarity

Unit #2

Unit #1

Load
Input

Input

Output

Ground

Output

Output

Ground

Output

COM

Note: The output reference ground (COMMON)
can be grounded at the power supply side instead
of the load, depending on the requirements. Local
sensing should be used in this configuration.

Caution

When connecting the units in series, diodes should
be connected across each output to prevent
reverse voltage.

Steps

1. Ensure the power is off on both power supplies.

2. Connect the master and slave unit in series as

shown above to either increase the voltage
output or to create a positive and negative
output. Remember that how the units are
grounded depends on the configuration of the
series connection.

3. Use diodes across the output terminals to

prevent reverse voltage at startup or if one of
the units unexpectedly shuts down. Ensure the
diodes are rated to withstand the voltage and
current output of the power supply.

4. Reattach the terminal cover.

Page 43