Grounding

The output terminals of the PSU power supplies are isolated with
respect to the protective grounding terminal. The insulation
capacity of the load, the load cables and other connected devices
must be taken into consideration when connected to the protective
ground or when floating.

Floating

As the output terminals are floating, the load
and all load cables must have an insulation
capacity that is greater than the isolation
voltage of the power supply.

PSU

Load

Ext-V

Ext-R

Analog

connector

( ) Insulation capacity

isolation voltage

of power supply

WARNING

If the insulation capacity of the load and load
cables are not greater than the isolation voltage of
the power supply, electric shock may occur.