ANALOG CONTROL

147

Connection- alt.
shielding

If the wire shield needs to be grounded at the
voltage source (EXT-V), then it can be
connected as shown below.

PSU

EXT-V

Isolated

Voltage

Option

4

3

Output

Terminal

2 core shielded

wire or twisted

pair

Pin4(GND_ISO) → EXT-V (-)

Pin3(+IPROG_ISO) → EXT-V (+)

Wire shield → EXT-V ground (GND)

Steps

1. Connect the external voltage according to the

connection diagrams above.

2. Set the F-91 power on

configuration setting to 4 (Control
by Isolated Board).

Page 111

3. Set the F-97 power on

configuration setting to 0 (0~5V
control range) or 1 (0~10V control
range).

Page 111

Be sure to cycle the power after the power
on configuration has been set.

4. Press the Function key and confirm

the new configuration settings (F-
91=4, F-97=0 or 1).

Function