APPENDIX

189

Analog Programming and Monitoring

Model

PSU 6-200

12.5-120

20-76 40-38 60-25

External voltage control
output voltage

Accuracy and linearity: ±0.5% of rated output
voltage.

External voltage control
output current

Accuracy and linearity: ±1% of rated output
current.

External resistor control
output voltage

Accuracy and linearity: ±1% of rated output
voltage.

External resistor control
output current

Accuracy and linearity: ±1.5% of rated output
current.

Output voltage monitor

Accuracy: ±1%

Output current monitor

Accuracy: ±1%

Shutdown control

Turns the output off with a LOW (0V to 0.5V)
or short-circuit.

Output on/off control

Possible logic selections:
Turn the output on using a LOW (0V to 0.5V)
or short-circuit, turn the output off using a
HIGH (4.5V to 5V) or open-circuit.
Turn the output on using a HIGH (4.5V to 5V)
or open-circuit, turn the output off using a
LOW (0V to 0.5V) or short-circuit.

Alarm clear control

Clear alarms with a LOW (0V to 0.5V) or short-
circuit.

CV/CC/ALM/PWR
ON/OUT ON indicator

Photocoupler open collector output;
Maximum voltage 30V, maximum sink current
8mA.

Trigger out

Maximum low level output = 0.8V; minimum
high level output = 2V; Maximum source
current = 8mA.

Trigger in

Maximum low level input voltage = 0.8V;
minimum high level input votage = 2.0V,
Maximum sink current = 8mA.