Set to C.C. Priority Mode

When setting the power supply to constant current mode, a voltage
limit must also be set to determine the crossover point. When the
voltage exceeds the crossover point, the mode switches to C.V.
mode. For details about C.C. operation, see page 23. C.C. and C.V.
mode have two selectable slew rates: High Speed Priority and Slew
Rate Priority. High Speed Priority will use the fastest slew rate for
the instrument while Slew Rate Priority will use a user-configured
slew rate.

Background

Before setting the power supply to
C.C. mode, ensure:

The output is off.

The load is connected.

Steps

1. Press the Function key. The

Function key will light up.

Function

2. The display will show the function (F-01) on the

voltage display and the setting for the function
in the current display.

VOLTAGE

CURRENT

LAN

RMT

ERR

DLY

ALM

ISR

M 1

M 2

M 3

RUN

C V

C C

V

A

Function

number

Function

setting

3. Rotate the Voltage knob to change

the F setting to F-03 (V-I Mode
Slew Rate Select).

Voltage