PSU Series User Manual

100

V-I Mode

Selects High Speed Priority or Slew Rate
Priority for CV or CC mode. The voltage or
current slew rate can only be edited if CC/CV
Slew Rate Priority is selected. The ISR indicator
will be lit for CC Slew Rate Priority and the
VSR indicator will be lit for CV Slew Rate
Priority.

Note: CC and CV Slew Rate Priority mode are
disabled when voltage/current output is set to
external control.

VOLTAGE

CURRENT

VSR

LAN

RMT

ERR

DLY

ALM

ISR

M 1

M 2

M 3

RUN

C V

C C

V

A

CC Slew Rate priority

VOLTAGE

CURRENT

VSR

LAN

RMT

ERR

DLY

ALM

ISR

M 1

M 2

M 3

RUN

C V

C C

V

A

CV Slew Rate Priority

F-03

0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority

Rising Voltage
Slew Rate

Sets the rising voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-04

0.001 ~ max. V/msec

Falling Voltage
Slew Rate

Sets the falling voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-05

0.001 ~ max. V/msec

Rising Current
Slew Rate

Sets the rising current slew rate. Only
applicable if V-I Mode is set to CC Slew Rate
Priority.
F-06

0.001 ~ max. A/msec