ANALOG CONTROL

129

Set F-94 (External output logic) in
the power on configuration
settings to 0 (High = On) or 1
(Low = On) and set F-98 (External
output control function) to 1(On).

Be sure to cycle the power after
setting the power on
configuration settings.

Page 111

2. Press the Function key and confirm

the new configuration settings (F-
94 = 0 or 1 and F-98=1).

Function

3. The switch is now ready to set the output on or

off.

Note

When using a switch over long distances, please
use a switch relay to extend the line from the coil
side of the relay.

Switch

Relay

Line

extention

Analog

connector

Output

Terminal

20

19

If a single switch control is to be used for multiple
units, please isolate each instrument. This can be
achieved by using a relay.

Warning

Ensure the cables used and the switch exceed the
isolation voltage of the power supply. For example:
insulation tubes with a withstand voltage higher
than the power supply can be used.