Psu series user manual, Slave units – GW Instek PSU-Series User Manual User Manual
Page 76
PSU Series User Manual
76
Note
Configuration settings can be checked on both the
master and slave units by pressing the Function key
and checking F-93.
Only the Master OVP, OCP and UVL settings are
used for protection. Slave protection levels are
disregarded.
OHP works independently for each unit.
Master-Slave
Operation
Only operate the power supplies in parallel if
the units are configured correctly.
Steps
1. Turn on the master and slave units. The slave
unit(s) will show a blank display.
Master unit
VOLTAGE
CURRENT
VSR
LAN
RMT
ERR
DLY
ALM
ISR
M 1
M 2
M 3
RUN
C V
C C
V
A
Slave units
VOLTAGE
CURRENT
VSR
LAN
RMT
ERR
DLY
ALM
ISR
M 1
M 2
M 3
RUN
C V
C C
V
A
2. Operation of all units is controlled
via the master unit. Operation of
the master unit is the same as for a
single unit. See the Basic Operation
chapter.
Page 51.
3. Press the Output key to
begin. The output LED will
become lit.
Output
Caution
Only operate the power supplies in parallel if using
units of the same model number.
Note
The panel controls are disabled on slave units,
including the output key. On slave units, only the
Function key can be used to view the current
settings.