OVP/OCP/ULV

OVP, OCP and ULV level for each unit must
be set separately.

The OVP and OCP protections are tripped
independently on the master and slave.

Remote monitoring

Voltage monitoring (VMON) and current
monitoring (IMON) should be performed on
both units.

The VMON voltage represents the voltage
of that particular unit.

Remote Sense

Please see the voltage sense chapter for
details, page 62.

External Voltage and Resistance Control

Voltage/Resistance controlled remote
control should be used on both units
separately.

The full scale voltage (in series) is equivalent
to the maximum external voltage or
resistance.

Slew Rate

The slave rate should be set for both units.

Internal Resistance

The internal resistance should be set for both
units.

Bleeder Control

The bleeder resistor setting should be set
equally on both units.